Technical Spezification
Main connection
Main voltage 230 V ~
Main frequency 50…60 Hz
Anschlusswert 1 kVA
The device can be supplied ex works for 110 V~ / 60 Hz.
Pulse shaping
Charger 500 J /s
Voltage adjustable (intensity) 15…30 kV
Capacitor 0,1 µF*
Magnetic field coil 4,75 µH*
Discharge, peak 2…4 A*
Resonant pulse frequency 200…230 kHz*
Modulated frequency 80 kHz…260 MHz
Pulse duration 100…500 µs
Flux density in the middle of the coil, peak 84 mT (400…800 G)
Repetition rate, adjustable1…3 Hz
Duration of administration, adjustable 5…15 min.
Insulation of the magnetic field coil
Nominal voltage of the insulation 30 kV
Test voltage of the insulation 65 kV DC 1 Minute
* (Some values are subject to tolerances of up ± 20% ab)